Infector Detector: Specifications

The system must be able to detect the presence of biofilms on urinary catheters by detection of AHL, at a minimum concentration of 5nM, and report with a visual signal within 3 hours. It must work within a temperature range of 20°-30°C, be portable and easy to use, have a shelf life of at least seven days, and must not be harmful or infectious.


Property
Value
Inputs
System must be sensitive to AHL concentration between 5-50nM
Outputs
System must give a visual signal if bacteria is present
Operating Conditions
System must operate within temperature 20-30°C
 
System must operate within pH range of 6-8
Health Regulations
System Must not be living replicating bacteria
Response Time
System needs to have a response time under 3 hour
Lifespan
System must have a shelf life of 7 days
Packaging
System must be portable and convenient to use


Specifications in detail

Input: AHL 5-50nM

Output: Fluorescent Protein

Operating Conditions

Health & Safety

Packaging and Shelf-life