Tokyo/Experimental precision and accurecy
From 2007.igem.org
(Difference between revisions)
(→Samples:) |
|||
(10 intermediate revisions not shown) | |||
Line 1: | Line 1: | ||
+ | __NOTOC__ | ||
== AHL assay== | == AHL assay== | ||
- | === | + | ===Objective: === |
To see how precisely we determine the fluorescence from GFP by removing LB media. | To see how precisely we determine the fluorescence from GFP by removing LB media. | ||
Line 15: | Line 16: | ||
<br>(100:0), (70:30), (50:50), (30:70), (10:90), (3:97), (1:99) | <br>(100:0), (70:30), (50:50), (30:70), (10:90), (3:97), (1:99) | ||
<br>[[Tokyo/Wash |Wash]] | <br>[[Tokyo/Wash |Wash]] | ||
- | ===Result: === | + | ===Result & conclusion: === |
- | = | + | In proportion to the percentages of GFP expressing cells, fluorescence was increased as shown in Fig. 1. |
+ | <br>This indicates that the "[[Tokyo/Wash |wash]]" process well removes the other factors influencing the fluorescence. | ||
+ | |||
+ | == == | ||
+ | [[Tokyo/Preliminary assays |"Back to Preliminary assays"]] |
Latest revision as of 18:48, 26 October 2007
AHL assay
Objective:
To see how precisely we determine the fluorescence from GFP by removing LB media.
Samples:
placQI + GFP in pTrc99A (pos. con.)
A4 ΔP+ GFP (neg. con.)
Procedure:
AHL assay Standard protocol
Make 200 ul of mixtures of pos. con. and neg. con.
The mixture ratios of pos. con. and neg. con. are
(100:0), (70:30), (50:50), (30:70), (10:90), (3:97), (1:99)
Wash
Result & conclusion:
In proportion to the percentages of GFP expressing cells, fluorescence was increased as shown in Fig. 1.